MICROMETRY
Micrometry is the measurement of microscopic objects using a calibrated eyepiece micrometer. The unit of measurement is the micron (u orum). The eyepiece micrometer scale requires calibration from a measured scale engraved on a micrometer glass slide (graticule).
Micrometry is the measurement of microscopic objects using a calibrated eyepiece micrometer. The unit of measurement is the micron (u orum). The eyepiece micrometer scale requires calibration from a measured scale engraved on a micrometer glass slide (graticule).
The eyepiece
micrometer is calibrated for each microscope objective at which the measurement
is required. For example, for parasitology, the scale should be calibrated to
40X objective while for haematology, it should be 100X objective.
Method of Calibrating an Eyepiece Micrometer
Example
Distance
measured
= 0.4 mm
Number of divisions
= 25 1
division measures 0.4/25 = 0.016 mm To convert mm toum 0.016 x 1000
1. Focus the
eyepiece micrometer scale and the object sharply.
2. Position
the object along the scale and count the divisions covered.
3. Refer to
the previously prepared table for the objective in use to obtain the
measurement for the number of divisions counted. This will give the size of the
objectin microns.
For example,
an object covering 2 divisions on the scale will measure 16 x 2 = 32 microns.
Table shows the relationship between units used for the measurement of
microorganisms.
Relationship between units used
for the measurement of microorganism,
|
|||
Unit of lenghth
|
Millimeter(mm)
|
Micrometer( UM)
|
Nanometer
|
Milimeter (mm)
|
1
|
1000(103)
|
1000.000 (106)
|
Micrometer(um)
|
0.0091 (103)
|
1
|
1000 (103)
|
Nanometer (nm)
|
0.000001 (106)
|
0.001(103)
|
1
|
Angstrom
|
0.0000001 (107)
|
0.0001(10-4)
|
0.1(10-1)
|
Method of Calibrating an Eyepiece
Micrometer
1. Unscrew
the upper lens of a 10X huygenian eyepiece and insert the eyepiece micrometer
disc, with engraved side facing down.
2. Place the
stage graticule slide on the stage; rotate the required objective into position
and bring the scale into focus. Both the eyepiece scale and the calibration
scale should be focused clearly.
3. Adjust
the field until O line of the eyepiece scale aligns exactly with the O line of
the calibration scale.
4. Look
along the scales and note where a division of the eyepiece scale exactly aligns
with a division of the calibration scale.
5. Measure
the distance between the 0 point and where the alignment occurs on the
graticule. The measurement of the calibration scale are 0.1 mm to 2.0 mm.Each
small subdivision measures 0.01 mm.
6. Count the
number of divisions of the eyepiece scale between the 0 point and where the
alignment occurs.
7. Calculate
the measurement for 1 division of the eyepiece scale in microns (um). Figure
shows calibration of eyepiece micrometer.